PELCON NT Build 492 Instrument

        The new PELCON NT Build 492 Instrument is designed & manufactured in DK from scientific grade components:

·         A tailored combination instrument with logical design and compact size

·         Precise voltage setting and easy operation by switching between cells

·         Output voltage: 0-60 Volts DC , output current: 0-350 mA per cell

·         Simultaneous control of 1-6 test cells

·         2 years warranty against false workmanship

  

 

Included accessories:

·         Housed safety cables

·         Unique set of stainless steel cathode/sample support & anode

·         Soft and durable rubber sleeves, Ø: 100(inner)×150mm

·         Heavy duty stainless steel clamps

·         Heavy duty catholyte reservoirs with hinged lids

·         Conditioning chamber

·         Honed digital slide caliper for precise ingress measurements

·         Digital thermometer

·         Spray bottle for the contrast liquid

 

 

 

     Peripherals offered:

    ·         Peripheral equipment, including Pelcon Vacuum Chamber and vacuum pump

    ·         Training course offered: installation of equipment, detailed testing procedures, pitfalls, calculation & evaluation of data;

relationship of mix design, microstructure and chloride migration coefficient

 

     Instrument specifications:

            Mains supply:                                                                                   230V ±10 %, 50 – 60 Hz

Power consumption:                                                                       10 – 180 W

Temperature range:                                                                         5 – 40 °C

Output voltage:                                                                                0 – 60 volts DC

Regulation at ±10 % change of mains voltage:                           ± 0.01 % typ.                                                       

Regulation at loads 0-100% (0…2A):                                            < 0,015%

Temperature coefficient at const. voltage:                                 < 0,005 %/°C typ.

Ripple and noise, 20 Hz-200 kHz:                                                 < 0.15mV rms.

Output current:                                                                               0-2A DC (0-350 mA per. cell)

Current measuring resistor:                                                          Precision foil power resistor 1R ± 0.1 % Type VPR221 Vishay precision group

Temperature coefficient of resistors:                                          ± 2ppm/°C typical

4 lead Kelvin connected device:                                                  Max ± 5 ppm, ref. +25°C                             

Voltage display:                                                                              4½ digit LCD digital panel meter

Performance:                                                                                  ±3 counts typ.

Temperature drift:                                                                         ± 0.6 counts/C typ.

Current display:                                                                             4½ digit LCD digital panel meter

Performance:                                                                                 ± 2 counts typ.

Temperature drift:                                                                        ± 0.4 counts/C typ.

Dimensions, W x D x H:                                                               323mm x 210mm x 80mm

Weight:                                                                                           5.5 kg

Finish:                                                                                             Polished stainless steel & blue enamel

 

 

  